CPAP watt-hour math only

Updated 9 September 2026

Not medical advice. Ranked does not recommend power stations for life-support or therapy. Jackery’s own 2000 v2 FAQ says the product is not recommended for medical devices or life-support. If a power interruption could cause harm, talk to a clinician and use equipment specified for that job. This page is arithmetic using a published power-supply spec, nothing else.

People still type “CPAP generator” into a search box. The only honest thing we can add is: look at the power-supply numbers, multiply by hours, and leave a margin. We are not telling you to run therapy from a camping battery.

Published numbers (ResMed AirSense 11 user guide)

ItemFigureSource
Power supply unit65 W classResMed AirSense 11 user guide
Typical power consumption56.1 W (111.5 VA)Same guide, 65 W PSU table
Peak power consumption73.2 W (137.6 VA)Same guide
DC output of PSU24 V, 2.71 ASame guide

Those “typical” and “peak” figures are manufacturer electrical specs for the 65 W supply, often with humidifier and heated tube in the test condition. Real draw with humidity off is often lower — we do not have a Ranked measurement, so we will not invent 20 W. If you need a number for math practice, use the published typical 56.1 W and treat it as conservative.

Arithmetic only

Energy per 8-hour night at 56.1 W: 56.1 × 8 = 448.8 Wh at the device. Through an 85% efficient AC inverter: 448.8 / 0.85 ≈ 528 Wh from the battery.

Nights at 56.1 W × 8 h, 85% inverterWh from batteryMfr. packs that exceed that Wh
1 night≈ 528 WhExplorer 500 v2 512 Wh is short. RIVER 2 Pro 768 Wh clears it. 1000 v2 1070 Wh clears it.
2 nights, no recharge≈ 1,056 Wh1000 v2 1070 Wh is on the edge. AC180 1152 Wh. 2000 v2 2042 Wh is comfortable on Wh alone.
3 nights, no recharge≈ 1,584 Wh2000 v2 2042 Wh. Not a 1 kWh sealed box.

Inverter watts: 73.2 W peak is far below every AC inverter on this site. The constraint is energy, not watts — if someone were using AC. A DC cable (where the manufacturer supports it) avoids inverter loss; we are not certifying any DC cable for a therapy device.

Again: this is not a recommendation to power a CPAP from a portable station. Manufacturer medical disclaimers exist for a reason.

Sources

ResMed AirSense 11 user guide (document.resmed.com), 65 W PSU table: typical 56.1 W, peak 73.2 W. Station watt-hours from manufacturer PDPs in README.md. Jackery 2000 v2 FAQ medical disclaimer on jackery.com. Checked 9 September 2026.